Skip to main content

Table 1 Mean values for D90, V100, V150 and V200 for the prostate along with the associated standard deviations

From: Assessment of I-125 seed implant accuracy when using the live-planning technique for low dose rate prostate brachytherapy

Dosimetry Parameter Planned St Dev Implanted St Dev
D90 (Gy) 175.5 9.6 177.7 10.5
V100 (%) 97.9 1.6 97.9 1.5
V150 (%) 64.1 7.3 66.5 7.9
V200 (%) 21.6 5.9 26.1 7.3