Table 2 Accuracy of isocenter relocation at CT verification

From: Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes

Direction (mm)

Mean

SD

Range

Cranio-caudal

0.4

0.6

-1,2 - 2.1

Medio-lateral

0.12

0.35

-0.9 - 1.4

Anterior-posterior

0.2

0.4

-1,6 - 1.3

3D-displacement

0.5

0.7

0.1-2.9