Skip to main content

Table 1 Treatment plan and delivery details for the different experiments.

From: Dosimetric precision of an ion beam tracking system

experiment

film response/cell samples

pin-point ionization chamber array

DDD

target volume [WxHxD mm3]

28 × 45 × 23

36 × 20 × 50

single beam

depth range [MeV/u]

265 - 299

267 - 330

400

number of iso-energy slices

18

21

-

number of raster points

3444

5946

1

grid spacing [mm]

2.5

2

-

scan path

up-down

up-down

-

beam spot FWHM [mm]

7.5

5.6

2.3

peak-to-peak amplitudelateral/radiological depth [mm/mm water]

40/15.2

30/11.4

40/15.2

motion period [s]

~ 3

~ 3.4

~ 3

target dose

6 GyE (3.3 Gy at target center)

2 Gy

-

irradiation time [s]

~ 140

~ 150

-