Skip to main content

Table 3 Random errors

From: Potentials of on-line repositioning based on implanted fiducial markers and electronic portal imaging in prostate cancer radiotherapy

 

Mean [mm] j = 1...23

Range [mm] j = 1...23

Random setup error

  

Left-right

2.3

-7.9/7.4

Superior-inferior

2.7

-8.6/7.8

Anterior-posterior

2.7

-12.1/5.4

Random prostate movement

  

Left-right

1.4

-5.5/6.3

Superior-inferior

2.3

-5.9/8.6

Anterior-posterior

2.7

-7.0/8.0

Random total error

  

Left-right

2.9

-8.0/6.6

Superior-inferior

3.9

-11.5/8.2

Anterior-posterior

4.3

-14.9/11.2

  1. The standard deviations of measured deviations Δsij, Δmij, Δtotij around the systematic errors (of each patient j = 1...23) <Δs>j, <Δm>j, <Δtot>j after averaging over i = 1,..., 8 weekly controls provide the random errors sqrt(1/7Σi = 1..8(<Δs>j - Δsij)2), sqrt(1/7Σi = 1..8(<Δmi>j - Δmij)2), sqrt(1/7Σi = 1..8(<Δtoti>j - Δtotij)2) for patient j = 1,..., 23. They are splitted into the RL, SI and AP direction for every error type, i.e. setup error, prostate movement and total error (see Fig. 1 for explanation of symbols). The mean random errors after averaging over 23 patients and their range, i.e. maximum deviations in both orientations of the axes, are given in the table.