Figure 5From: Functional intercomparison of intraoperative radiotherapy equipment – Photon Radiosurgery SystemOutput Diminution Factor. To determine the constancy of output from the x-ray sources, the ion chamber current was monitored over a period of 30 minutes equivalent to typical clinical treatment duration. A difference between the initial and final current readings was consistently observed and the mean reduction was 1.58%.Back to article page