Figure 3 | Radiation Oncology

Figure 3

From: Phantom investigation of 3D motion-dependent volume aliasing during CT simulation for radiation therapy planning

Figure 3

Magnitude of mis-estimation of the true physical target volumes (tTV) of the 10- and 31.5-mm diameter targets as a function of eight (8) initial motion phases for three (3) motion amplitudes. The mis-estimation magnitude is computed as a ratio of each CT reconstructed dGTV and its corresponding tTV. Plots A, B, and C correspond to single-slice fast (or 1-sec), multi-slice fast (or 1-sec), and single-slice slow (or 4-sec) scan imaging techniques, respectively. Each colored line represents a specific motion amplitude in the S-I direction, synchronized with constant amplitudes of ± 2 and ± 5 mm in the R-L and A-P, respectively.