Table 2 Additional output factors at SSD 80 cm, 10 cm depth

From: MR compatible detectors assessment for a 0.35 T MR-linac commissioning

Field size (cm2)

Mean measured OF

Relative standard deviation (%)

Mean corrected OF

Relative standard deviation

0.84 × 0.83

0.643

2.0

0.628

0.5

1.66 × 1.66

0.787

1.0

0.785

0.4

2.5 × 2.49

0.828

0.8

0.831

0.4

3.32 × 3.32

0.858

0.7

0.861

0.4

4.16 × 4.15

0.885

0.6

0.888

0.3

6.64 × 6.64

0.945

0.3

0.947

0.2

8.3 × 8.3

0.975

0.2

0.976

0.1

9.96 × 9.96

1.000

0.0

1.000

0.0