Table 1 Characteristics of the detectors and their use in the study

From: MR compatible detectors assessment for a 0.35 T MR-linac commissioning

 

Detectors

Collecting volume

Constructor field size range of use

Type of measurement performed

Ionization chambers

Exradin A28 MR

0.125 cc

3–40 cm2

Output calibration

PTW Semiflex 31010

0.125 cc

3–40 cm2

Output calibration/PDD/Profiles/Output factors

PTW Semiflex 3D 31021

0.07 cc

2.5–4 cm2

Output calibration/PDD/Profiles/Output factors

PTW Markus 34045

0.02 cc

3–40 cm2

PDD

PTW Pinpoint 3D 31022

0.016 cc

0.8–40 cm2

PDD/Profiles/Output factors

High resolution detectors

Sun Nuclear Edge (Shielded diode)

0.019 mm3

0.5–10 cm2

PDD/Profiles/Output factors

PTW 60019 microDiamond

0.004 mm3

0.4–40 cm2

PDD/Profiles/Output factors

PTW 60023 microSilicon (unshielded diode)

0.03 mm3

0.4–10 cm2

PDD/Profiles/Output factors

Passive detectors

EBT3 radiochromic films

N/A

N/A

Output factors

Lif microcubes

1 mm3

N/A

Output factors