Table 1 Overview of verifications throughout the treatment workflow. In case a verification fails, the patient is transferred from the MRI-only workflow to the standard workflow based on CT

From: “sCT-Feasibility” - a feasibility study for deep learning-based MRI-only brain radiotherapy

No.

Verification criterion

1

Can the sCT be generated and is the sCT clinically utilizable?

2

Are the three rotations needed for CT-MRI registration each ≤ 3°?

3

Can a treatment plan be generated and verified using the sCT?

4

Is the dosimetric difference between sCT- and CT-based treatment plans in the

PTV ≤ 3%?

5

Is the dosimetric difference between the sCT- and CT-based treatment plans in affected OARs (receiving > 10% of prescribed dose) ≤ 3%?

6

Are the couch correction parameters during patient positioning in the rotational degrees of freedom ≤ 3°?