Skip to main content

Table 3 Summary of volunteer results

From: Repeatability quantification of brain diffusion-weighted imaging for future clinical implementation at a low-field MR-linac

ROI

highRes

highSNR

Literature

Scan 1

Scan 2

RC

relRC

Scan 1

Scan 2

RC

relRC

 

Unit

10–6 mm2/s

10–6 mm2/s

10–6 mm2/s

%

10–6 mm2/s

10–6 mm2/s

10–6 mm2/s

%

10–6 mm2/s

CSF

2264 ± 484

2258 ± 428

167 [117, 293]

7 [5, 12]

2868 ± 624

2878 ± 729

364 [254, 639]

12 [9, 22]

2730–3020

ROIleft

528 ± 104

532 ± 107

62 [43, 108]

12 [8, 20]

680 ± 82

676 ± 82

41 [28, 71]

6 [4, 10]

620–1090

ROIright

518 ± 105

515 ± 106

66 [46, 116]

13 [9, 23]

673 ± 81

672 ± 80

46 [32, 80]

7 [5, 12]

620–1090

ROIpost

668 ± 144

686 ± 139

70 [49, 123]

11 [8, 19]

816 ± 117

822 ± 122

64 [45, 112]

8 [5, 14]

620–1090

ROIant

750 ± 123

748 ± 128

34 [24, 60]

5 [3, 8]

798 ± 92

804 ± 104

41 [29, 72]

5 [4, 9]

620–1090

All 4 cyl. ROIs

  

53 [44, 68]

9 [8, 12]

  

44 [36, 56]

6 [5, 7]

 
  1. The ADCs (mean ± 1σ) within each ROI, averaged over all ten volunteers are reported for both sequence variants and scans, alongside literature values [38]. Additionally, the RC and relRC values with their respective 95% confidence intervals are reported for each ROI separately, and for all four cylindrical ROIs combined