Skip to main content

Table 1 Parameters of the two investigated EPI DWI sequence variants

From: Repeatability quantification of brain diffusion-weighted imaging for future clinical implementation at a low-field MR-linac

Sequence variant

High resolution (“highRes”)

High SNR (“highSNR”)

b-values (s/mm2)

0, 100, 250, 500, 800

0, 100, 250, 500, 800

Acquisition time (min:s)

6:36

6:20

Slice thickness (mm)

5

7

Number of slices

20

14

Acquisition matrix

90 × 90

80 × 80

In-plane acquisition matrix resolution (mm2)

3.0 × 3.0

3.5 × 3.5

In-plane image resolution after zero-filling interpolation (mm2)

1.5 × 1.5

1.75 × 1.75

Field-of-view (mm3)

270 × 270 × 100

280 × 280 × 98

Number of averages

9

11

TR/TE (ms)

3300/110

2600/110

Bandwidth (Hz/pixel)

1355

1359

Flip angle

90°

90°