Table 2 Comparison of IC doses in four algorithms with the measeured doses

From: Evaluation of heterogeneity dose distributions for Stereotactic Radiotherapy (SRT): comparison of commercially available Monte Carlo dose calculation with other algorithms

 

Pinpoint chamber

SHM

error

CCC

error

SP

error

MC

error

 

[cGy]

[cGy]

[%]

[cGy]

[%]

[cGy]

[%]

[cGy]

[%]

Single Beam (3 × 3 cm2) in Water Phantoma)

124.6

124.0

0.48

124.4

0.16

125.5

-0.72

123.9

0.56

Eight Beams (4 × 4 cm2) in inhomogeneous Phantomb)

200.9

209.6

-4.33

199.9

0.50

200.6

0.15

201.7

-0.40

  1. Abbreviations: SHM, scatter homogeneous mode, CCC, C.C.Convolution, SP, Superposition, MC, Monte Carlo.
  2. a) Single delivery of 200MU from gantry angle of 0 degree.
  3. b) Static 8 beams; the gantry, couch angles and MU were 180° + 0° 33.5MU, 260° + 0° 44.3MU, 340° + 0° 33.6MU, 30° + 40° 34.1MU, 35° + 320° 35.3MU, 320° + 320° 36.1MU, 30° + 90° 34.4MU, 330° + 90° 34.5MU, respectively. The MU employed here has equal weights in IC dose of CCC in inhomogeneous phantom.