Skip to main content

Table 2 Summary of dose volume histogram analysis for PTV and healthy tissue

From: Feasibility and early clinical assessment of flattening filter free (FFF) based stereotactic body radiotherapy (SBRT) treatments

Parameter

All

Lung

Liver

Abdomen

PTV

Volume [cm 3 ]

78 ± 78

55 ± 55

146 ± 116

115 ± 82

Mean dose [%]

101.0 ± 2.9

101.2 ± 1.0

100.6 ± 1.6

100.3 ± 0.5

St. Dev. [%]

2.7 ± 0.6

2.9 ± 0.5

2.3 ± 0.8

2.1 ± 0.6

D 1% [%]

106.0 ± 1.8

106.8 ± 1.5

104.3 ± 1.6

104.4 ± 1.4

D 95% [%]

96.2 ± 1.0

96.1 ± 0.9

96.2 ± 1.5

96.4 ± 0.8

V 95% [%]

97.3 ± 1.9

97.1 ± 1.7

97.2 ± 3.0

97.9 ± 1.5

V 107 [%]

1.5 ± 5.1

2.1 ± 6.0

0.1 ± 0.2

0.0 ± 0.1

Healthy tissue

Volume [cm 3 ]

30660 ± 8928

29283 ± 7493

36232 ± 13301

31510 ± 8805

Mean [Gy]

1.2 ± 0.7

1.1 ± 0.8

1.4 ± 0.3

1.3 ± 0.6

V 10 Gy [%]

3.3 ± 2.7

3.0 ± 3.0

4.0 ± 1.3

3.9 ± 2.4

CI 95%

1.2 ± 0.3

1.3 ± 0.3

1.2 ± 0.3

1.2 ± 0.3

Dose Int. [Gy*cm 3 *10 4 ]

3.5 ± 2.1

3.1 ± 2.0

5.0 ± 2.1

4.1 ± 1.8

  1. V xGy : volume receiving at least × Gy. D x% : dose received by at least x% of the volume. CI: Conformity Index