Table 3 Radial dose function calculated for the new GZP3 60Co afterloading system by Geant4 and EGSnrc

From: Monte Carlo dosimetry of the 60Co sources of a new GZP3 HDR afterloading system

r (cm)

g(r)

Channel 1 and 2

Channel 3

Geant4

EGSnrc

Geant4

EGSnrc

0.25

0.963

0.965

0.964

0.969

0.5

1.037

1.035

1.033

1.039

0.75

1.015

1.010

1.011

1.015

1

1.000

1.000

1.000

1.000

1.5

0.991

0.991

0.989

0.994

2

0.981

0.982

0.981

0.985

2.5

0.971

0.973

0.972

0.976

3

0.966

0.965

0.963

0.967

4

0.950

0.947

0.947

0.950

5

0.936

0.931

0.935

0.933

6

0.917

0.914

0.921

0.916

7

0.898

0.898

0.901

0.900

8

0.883

0.882

0.885

0.884

10

0.852

0.850

0.849

0.853