From: Monte Carlo dosimetry of the 60Co sources of a new GZP3 HDR afterloading system
Source design | Λ (cGy h−1 U−1) |
---|---|
New GZP3 60Co Channel 1 and 2 (This work by Geant4) | 1.115 ± 0.004 |
New GZP3 60Co Channel 1 and 2 (This work by EGSnrc) | 1.112 ± 0.004 |
New GZP3 60Co Channel 3 (This work by Geant4) | 1.116 ± 0.004 |
New GZP3 60Co Channel 3 (This work by EGSnrc) | 1.113 ± 0.004 |
GZP6 60Co (MCNP) [8] | 1.104 ± 0.003 |
BEBIG 60Co (Geant4) [24] | 1.084 ± 0.005 |
Flexisource 60Co (Geant4) [26] | 1.085 ± 0.003 |