Table 1 Accuracy results of CLOD according to the compensating material in the transmission and reflective modes

From: Scanning methodology for contact lens-type ocular in vivo dosimeter (CLOD) dosimetry applying a silicone material

Delivered dose

25 cGy

70 cGy

Scan mode

Transmission mode

Reflective mode

Transmission mode

Reflective mode

Materials

Dose (cGy)

Dose diff. (%)

Dose (cGy)

Dose diff. (%)

Dose (cGy)

Dose diff. (%)

Dose (cGy)

Dose diff. (%)

Air

26.7

6.8

26.3

5.2

74.6

6.6

74.1

5.9

DS

26.4

5.4

23.8

4.8

74.5

6.4

73.9

5.5

SC

23.7

5.4

26.2

4.7

73.6

5.2

73.4

4.9

OG

25.6

2.2

25.3

1.3

70.2

0.3

70.1

0.2