Skip to main content

Table 2 Average difference of each time component compared with log files

From: Developing an accurate model of spot-scanning treatment delivery time and sequence for a compact superconducting synchrocyclotron proton therapy system

Time component

Relative difference in percentage

Absolute difference (s)

ELST

(− 3.1 ± 5.7)% ([− 8.8%, 2.6%])

 − 0.8 ± 1.4

SSWT

(5.9 ± 3.9)% ([2.0%, 9.8%])

3.1 ± 2.9

SSPT

(2.6 ± 8.7)% ([− 6.1%, 11.3%])

0.2 ± 1.8

BST

(− 2.3 ± 5.3)% ([− 7.6%, 3.0%])

 − 0.5 ± 0.9

Total BDT

(2.1% ± 3.0)% ([− 0.9%, 5.1%])

2.1 ± 3.6