Fig. 6 | Radiation Oncology

Fig. 6

From: How can we consider variable RBE and LETd prediction during clinical practice? A pediatric case report at the Normandy Proton Therapy Centre using an independent dose engine

Fig. 6

Evaluation of VSAD implementation and impact on GPU-accelerated ray tracing and dose kernel execution via beam projection (spot position and FWHM) evaluation of FRoG against reference RS-MC calculation for the 16-spot plan. The lower and upper limit of the commissioned clinical energy range (98 MeV and 226 MeV) are tested to demonstrate depth and energy dependence of beam propagation. X and Y line profiles are presented for entrance channel (EC) and Bragg peak (BP)