From: Feasibility study of volumetric modulated arc therapy with Halcyon™ linac for total body irradiation
Segment | Seg 1 | Seg 2 | Seg 3 | Seg 4 | Seg 5 | Seg 6 | Seg 7 |
---|---|---|---|---|---|---|---|
EPID | |||||||
1%/1 mm | 96.6 | 96.2 | 96.3 | 97.2 | 97.3 | 97.3 | 98.0 |
2%/2 mm | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
ArcCHECK | |||||||
2%/2 mm | 91.7 | 89.9 | 93.6 | 81.0 | 81.6 | 84.4 | 90.6 |
3%/2 mm | 97.7 | 97.8 | 99.4 | 93.6 | 94.6 | 93.8 | 96.8 |
3%/3 mm | 98.2 | 98.9 | 99.7 | 96.3 | 97.1 | 97.0 | 98.9 |
Point dose discrepancy (%) | 0.60 | 0.70 | 2.48 | 2.16 | 0.20 | 1.59 | 0.43 |