Table 1 Measured parameters and values set based on recommendations and intended practices

From: Analytical setup margin for spinal stereotactic body radiotherapy based on measured errors

Penumbral width

3.57 mm

3.57 mm

Systematic error

0.611 mm

0.611 mm

Random error

1.31 mm

1.31 mm

Fractions

5

1

Population percentage

90%

90%

Dose percentage

90%

90%

Resulting margin

2.42 mm

3.61 mm

  1. These values were input into the margin recipe and will be used as the standard values with which to compare for sensitivity testing