Table 2 DVH comparison between Synergy and Halcyon plans

From: Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems

 

Synergy Agility

Halcyon

p value

PTV

   

D95 in Gy

33.9 ± 2.3

33.5 ± 2.5

NS*

D80 in Gy

38.0 ± 0.5

38.6 ± 0.4

p < 0.01

DMax in Gy

43.0 ± 0.6

43.3 ± 0.8

NS

HI

0.13 ± 0.01

0.12 ± 0.02

NS

PTV-0.5

   

DMin in Gy

35.3 ± 1.0

35.4 ± 0.2

NS

D95 in Gy

38.5 ± 0.2

38.8 ± 0.3

0.03

DMean in Gy

40.1 ± 0.1

40.1 ± 0.1

NS

SD in Gy

1.0 ± 0.1

0.8 ± 0.1

0.01

Lung ipsilateral

   

DMean in Gy

4.7 ± 0.9

4.4 ± 1.1

NS

Lung contralateral

   

DMean in Gy

0.3 ± 0.1

0.2 ± 0.1

p < 0.01

Breast contralateral

   

DMean in Gy

0.5 ± 0.2

0.3 ± 0.1

p < 0.01

D01 in Gy

1.8 ± 0.7

1.0 ± 0.5

p < 0.01

Heart**

   

DMax in Gy

31.7 ± 10.5

33.0 ± 8.4

NS

DMean in Gy

1.7 ± 0.4

1.5 ± 0.5

p < 0.01

  1. Dose to target volumes and organs at risk in Synergy Agility plans and Halcyon plans. Stated values indicate mean ± SD and statistical analysis was performed using the Wilcoxon test. The differences were considered statistically significant when p < 0.05
  2. *Not significant (NS)
  3. **Only left-sided plans were considered for the evaluation of the heart dose