Table 4 Comparison of standard deviations (SD) in mm of intra-fraction positioning errors found in the literature as well as the Pos_Old and Pos_New positioning systems

From: Evaluation of inter- and intra-fraction 6D motion for stereotactic body radiation therapy of spinal metastases: influence of treatment time

Authors

Positioning systems

SD (mm) for intra-fraction CBCT

Hyde et al

Elekta BodyFIX system

LR = 0.6; CC = 0.5 and AP = 0.5

Our center

Pos_New

LR = 0.5; CC = 0.5 and AP = 0.7

Dahele et al

Posirest

LR = 0.9; CC = 0.6 and AP = 0.7

Li et al

Elekta BodyFIX system

LR = 0.9, CC = 0.7 and AP = 0.9

Li et al

Evacuated cushion

LR = 1.3; CC = 1.2 and AP = 1.0

Li et al

Thermoplastic S-frame

LR = 1.3, CC = 0.9, and AP = 1.1

Foster et al

Evacuated cushion

LR = 1.05; CC = 1.23 and AP = 1.04

Our center

Pos_Old

LR = 0,9; CC = 1,5 and AP = 0,7