Skip to main content

Table 1 Plan quality evaluation criteria. The planning goals for tumor target and dose constraints for OARs were listed

From: Evaluation of a highly refined prediction model in knowledge-based volumetric modulated arc therapy planning for cervical cancer

 

Criteria

Quality scores

Acceptable

Excellent

PGTVnd

V60Gy (%) ≥ 97

V60Gy (%) = 100

4.8/8

 

V66Gy (%) ≤ 10

V66Gy (%) ≤ 5

4.8/8

 

Dmin (Gy) ≥ 54.6

Dmin (Gy) ≥ 55.8

3/5

 

CI (0–1):

As high as possible

0–2

 

HI (0–1):

As low as possible

0–2

PCTV

V45Gy (%) ≥ 97

V45Gy (%) = 100

4.8/8

 

V49.5 Gy (%) ≤ 20

V49.5 Gy (%) ≤ 10

4.8/8

 

Dmin (Gy) ≥ 40.95

Dmin (Gy) ≥ 41.85

3/5

 

CI (0–1):

As high as possible

0–2

 

HI (0–1):

As low as possible

0–2

Spinal cord

D0.03 cc (Gy) ≤ 45

D0.03 cc (Gy) ≤ 40

6/10

Bladder

D35.0% (Gy) ≤ 50

D35.0% (Gy) ≤ 45

6/10

Rectum

D60.0%(Gy) ≤ 45

D60.0%(Gy) ≤ 40

6/10

Kidney (L)

V18Gy (%) ≤ 32

V18Gy (%) ≤ 20

3/5

Kidney (R)

V18Gy (%) ≤ 32

V18Gy (%) ≤ 20

3/5

Femoral head (L)

V35Gy (%) ≤ 50

V35Gy (%) ≤ 15

1.8/3

 

D0.03 cc (Gy) ≤ 65

D0.03 cc (Gy) ≤ 50

1.2/2

Femoral head (R)

V35Gy (%) ≤ 50

V35Gy (%) ≤ 15

1.8/3

 

D0.03 cc (Gy) ≤ 65

D0.03 cc (Gy) ≤ 50

1.2/2

Total score

/

/

100