Skip to main content

Table 3 Plan parameter comparison between vIMPT and SPArc

From: Feasibility study: spot-scanning proton arc therapy (SPArc) for left-sided whole breast radiotherapy

Plan parameters

vIMPT

SPArc

Absolute difference (SPArc − vIMPT)

Beam directions

1

39

38

Total energy layers

27 ± 3.85

93 ± 4.57

66 ± 6.95

Total monitor unit

6143 ± 1281.08

5511 ± 1233.95

− 633 ± 140.91

Total delivery time (5 s)

758 ± 144.17

1059 ± 123.77

301 ± 30.34 (p < 0.001)

Total delivery time (4 s)

732 ± 141.3

967 ± 126.08

235 ± 24.17 (p < 0.001)

Total delivery time (3 s)

706 ± 138.53

874 ± 128.50

169 ± 18.41 (p < 0.001)

Total delivery time (2 s)

680 ± 135.79

782 ± 131.04

102 ± 13.64 (p < 0.001)

Total delivery time (1 s)

654 ± 133.11

690 ± 133.68

36 ± 11.18 (p < 0.001)

Total delivery time (0.5 s)

641 ± 131.79

644 ± 135.03

3 ± 11.35 (p = 0.47)

Total delivery time (0.2 s)

633 ± 131.00

616 ± 135.86

− 17 ± 11.93 (p = 0.005)