Skip to main content

Table 3 Comparison of planned and ionization chamber measured point dose at five representative points of five anatomical sites of each TMI/TMLI patient

From: Leaf open time sinogram (LOTS): a novel approach for patient specific quality assurance of total marrow irradiation

Site P11 P12 P13 P14 P15
Planned (cGy) Measured (cGy) Planned (cGy) Measured (cGy) Planned (cGy) Measured (cGy) Planned (cGy) Measured (cGy) Planned (cGy) Measured (cGy)
Brain 164 163.91 165 166.47 162 163.12 164 162.09 186 185.66
Chest 177 177.02 193 193.86 198 198.87 169 168.5 184 182.69
Pelvis 191 190.77 205 205.31 201 202.23 174 173.5 208 210.34
Upper leg 149 149.34 172 172.54 151 151.33 152 151.69 169 167.27
Lower leg 141 141.26 153 153.42 149 149.48 149 148.03 186 185.77