Table 1 Summary of four cone-beam CT imaging protocols used in this study

From: Image quality of 4D in-treatment CBCT acquired during lung SBRT using FFF beam: a phantom study

 

3D

4D

Pretreatment

In-treatment

Pretreatment

In-treatment

Image acquisition parameters

 MV treatment beam

No

Yes

No

Yes

 Tube voltage (kV)

120

120

120

120

 Tube current (mA)

40

40

20

20

 Exposure time (ms)

40

40

16

16

 Collimator

M20

M20

S20

S20

 Rotation angle (°)

360

360

200

200

 Gantry speed (°/min)

360

67

 Acquisition interval (°)

0.1

0.1

 Estimated time (s)a

60

99

180

55

Analysis results

 #projections acquired

333

512

1010

266

 Contrast (%)

9.6

10.1

8.0

5.5

 Contrast-to-noise ratio

10.0

9.5

10.7

2.6

 Overall uniformity (%)

98.5

98.0

97.1

89.1

 Minimum uniformity (%)

98.8

98.4

98.4

95.7

  1. The CBCT acquired without MV treatment beam delivery is denoted by pretreatment CBCT as it is acquired prior to treatment. The results of an initial image quality test for the listed protocols are summarized
  2. aThe time for the in-treatment CBCT images was estimated as the beam-on time expected to deliver the VMAT plan for patient 1. In general, the time to acquire an in-treatment CBCT image varies depending on the VMAT plan, which is concurrently delivered