Skip to main content

Table 5 Principal component analysis of the placement variations of the transducer arrays: Placement variations along the principal components. Kolmogorov-Smirnov test: Test for comparison of the placement deviations from the overall mean per patient with a normal distribution

From: Combined radiotherapy and concurrent tumor treating fields (TTFields) for glioblastoma: Dosimetric consequences on non-coplanar IMRT as initial results from a phase I trial

Array

First principal component

Second principal component

Third principal component

Standard deviation [cm]

p-value

Kolmogorov-Smirnov test

Standard deviation [cm]

p-value

Kolmogorov-Smirnov test

Standard deviation [cm]

p-value

Kolmogorov-Smirnov test

frontal

1.06

< 0.01

0.70

< 0.01

0.31

< 0.01

occipital

1.23

0.01

0.71

< 0.01

0.25

< 0.01

left

0.96

> 0.15

0.79

> 0.15

0.28

< 0.01

right

1.11

> 0.15

0.68

0.02

0.25

> 0.15