Skip to main content

Table 3 Global gamma pass-rates for comparison between sCT and CT dose-distributions using a 15% dose cut-off

From: MR-PROTECT: Clinical feasibility of a prostate MRI-only radiotherapy treatment workflow and investigation of acceptance criteria

Gamma criteriaGamma pass rate ± 1 SD [range] (%)
3%/3 mm99.8 ± 0.2 [99.2–100]
3%/2 mm99.7 ± 0.3 [98.9–100]
2%/2 mm99.7 ± 0.3 [98.7–100]
2%/1 mm99.1 ± 0.4 [98.0–99.7]