Skip to main content

Table 1 Worst and mean observed distortion measured from March 2017 to May 2018 on the MR-scanner

From: MR-PROTECT: Clinical feasibility of a prostate MRI-only radiotherapy treatment workflow and investigation of acceptance criteria

Geometric distortion over 15 months (mm)

Radial distance from isocenter

< 100

100–150

150–200

200–250

Mean distortion (1 SD) [range]

0.2 (0.0) [0.1–0.2]

0.3 (0.0) [0.2–0.4]

0.5 (0.1) [0.4–0.7]

1.9 (0.1) [1.7–2.0]

Mean of max distortion (1 SD) [range]

0.6 (0.1) [0.4–0.7]

0.8 (0.2) [0.6–1.1]

1.6 (0.3) [1.4–2.6]

7.9 (0.3) [7.4–8.6]