Skip to main content

Table 1 Percentage of PDD dose points passing the three Γ-criteria. for varying histories at an initial beam energy of 6.26 MeV

From: Linac photon beam fine-tuning in PRIMO using the gamma-index analysis toolkit

Number of Histories (106)

Average Dose Uncertainty (%)

3 % /3mm (%)

2 % /2mm (%)

1 % /1mm (%)

2

36.4

91.2

57.5

17.5

5

18.0

97.4

74.0

24.7

10

12.1

100.0

88.0

38.0

25

7.9

100.0

99.7

74.7

100

4.1

100.0

100.0

84.1