Skip to main content

Table 2 Contrast-noise ratio (CNR) for different fiducial markers

From: Fiducial markers visibility and artefacts in prostate cancer radiotherapy multi-modality imaging

Planar image

RP

CIVCO

GA

GA (B)

PM

EPID integrated

6MV 200MU/min ET 3379 s

 @0o

8.36

15.48

2.61

1.67

nv

 @90o

9.73

4.93

nv

3.18

nv

6MV

6MV MsE 3 MU

 @0o

6.04

8.33

nv

nv

nv

 @90o

4.44

5.60

nv

nv

nv

2.5MV

2.5MV MsE 1.5 MU

 @0o

8.56

7.39

4.24

5.94

nv

 @90o

4.15

4.51

2.63

3.59

nv

kV

0o: kVp 85,134 mA ET 112 s

90o: kVp 110,102 mA ET 147 s

 @0o

2.64

2.09

4.82

6.40

9.38

 @90o

6.08

6.40

4.62

3.95

2.35

ExacTrac

212o: kVp 100,400 mA ET 100 s

149o: kVp 100,400 mA ET 100 s

 Tube1

2.78

3.42

2.57

2.57

3.17

Tube2

2.60

2.47

3.21

3.90

2.97

  1. MsE MeterSet Exposure, nv not visible, italic poor visibility