Table 4 Differences in mechanical parameters between machine log files and original volumetric modulated arc therapy (VMAT) plans

From: Prediction of VMAT delivery accuracy with textural features calculated from fluence maps

Treatment site

MLC error (mm)

Gantry angle error (°)

Monitor unit error (MU)

Trilogy

 H&N

0.188 ± 0.065

0.054 ± 0.002

0.104 ± 0.072

 Prostate (PP)

0.043 ± 0.082

0.053 ± 0.004

0.061 ± 0.021

 Prostate (BP)

0.027 ± 0.011

0.054 ± 0.002

0.055 ± 0.011

 Liver

0.122 ± 0.098

0.054 ± 0.002

0.076 ± 0.037

 Spine

0.073 ± 0.061

0.054 ± 0.001

0.124 ± 0.042

TrueBeam STx

 H&N

0.090 ± 0.012

0.029 ± 0.000

0.017 ± 0.005

 Brain

0.040 ± 0.022

0.029 ± 0.002

0.052 ± 0.041

 Lung (SABR)

0.011 ± 0.006

0.010 ± 0.001

0.396 ± 0.033

 Spine (SABR)

0.024 ± 0.011

0.025 ± 0.003

0.257 ± 0.088

 Liver (SABR)

0.029 ± 0.018

0.037 ± 0.025

0.189 ± 0.106

  1. MLC Multi-leaf collimator, H&N head and neck, PP primary plan, BP boost plan, SABR stereotactic ablative radiotherapy