Skip to main content

Table 2 The 5-arc plan template for LINAC larynx SBRT planning

From: Benchmarking techniques for stereotactic body radiotherapy for early-stage glottic laryngeal cancer: LINAC-based non-coplanar VMAT vs. Cyberknife planning

Field

Energy

Gantry Rotation

Collimator Angle

Couch Angle

 Arc 1

6x

270° CW 179°

10°

10°

 Arc 2

6x

90° CCW 181°

350°

350°

 Arc 3

6x

330° CW 20°

10°

90°

 Arc 4

6x

330° CW 20°

10°

90°

 Arc 5

6x

179° CCW 181°

350°

0°