Table 3 Output Factors. Column 2 shows commissioning quality as the Precision MC commissioning process fit OFs to measured

From: Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments

Field Size [mm] (Commissioning fit - measured) / measured [%] (Precision MC - measured) / measured [%] (IDC – measured) / measured [%] (Precision MC - IDC) / IDC [%]
7.6 × 7.7 0.1 − 1.9 −4.3 2.5
15.4 × 15.4 − 0.1 − 1.4 − 1.1 − 0.3
23.0 × 23.1 −0.1 −1.0 − 0.5 −0.5
30.8 × 30.8 −0.1 −1.2 − 0.1 − 1.0
38.4 × 38.4 0.1 −1.0 0.0 −1.0
46.2 × 46.2 0.0 −1.2 −0.3 − 0.9
53.8 × 53.9 0.0 −2.0 −0.2 −1.7
69.2 × 69.3 0.1 −1.2 −0.5 − 0.6
84.6 × 84.7 0.2 0.5 −0.2 0.7
100.0 × 100.1 0.1 − 0.8 −0.2 − 0.7
115.0 × 100.1 0.0 0.2 −0.3 0.4