Skip to main content

Table 3 Output Factors. Column 2 shows commissioning quality as the Precision MC commissioning process fit OFs to measured

From: Benchmarking Monte-Carlo dose calculation for MLC CyberKnife treatments

Field Size [mm]

(Commissioning fit - measured) / measured [%]

(Precision MC - measured) / measured [%]

(IDC – measured) / measured [%]

(Precision MC - IDC) / IDC [%]

7.6 × 7.7

0.1

− 1.9

−4.3

2.5

15.4 × 15.4

− 0.1

− 1.4

− 1.1

− 0.3

23.0 × 23.1

−0.1

−1.0

− 0.5

−0.5

30.8 × 30.8

−0.1

−1.2

− 0.1

− 1.0

38.4 × 38.4

0.1

−1.0

0.0

−1.0

46.2 × 46.2

0.0

−1.2

−0.3

− 0.9

53.8 × 53.9

0.0

−2.0

−0.2

−1.7

69.2 × 69.3

0.1

−1.2

−0.5

− 0.6

84.6 × 84.7

0.2

0.5

−0.2

0.7

100.0 × 100.1

0.1

− 0.8

−0.2

− 0.7

115.0 × 100.1

0.0

0.2

−0.3

0.4