Table 5 Calculated mean NTCP values (%) of all patients

From: A treatment planning study comparing IMRT techniques and cyber knife for stereotactic body radiotherapy of low-risk prostate carcinoma

Technique/Organ Rectum Urinary bladder
HT 2.3* 2.8
SW 3.8 1.8
VMAT 4.5 2.8
CK 5.8 2.2
  1. Abbreviations: HT Helical tomotherapy, SW Sliding Window, VMAT Volumetric Modulated Arc Therapy, CK Cyber Knife; * - value with statistically significant difference (p < 0.05)