Skip to main content

Table 1 The dose-volume constraints

From: Optimal dose limitation strategy for bone marrow sparing in intensity-modulated radiotherapy of cervical cancer

Structure

Dose-volume constraints

Relative Priority

PTVo

D100% ≥ 50.70 Gy

150

Dmax ≤ 53.50 Gy

150

PTVi

D100% ≥ 50.70 Gy

150

Dmax ≤ 51.50 Gy

150

BM

Dmean ≤ 32.00 Gy

80

OC

Dmean ≤ 28.00 Gy

80

LS

Dmean ≤ 35.00 Gy

80

Small bowel

Dmax ≤ 52.50 Gy

80

V40 < 30%

80

Rectum

Dmax ≤ 52.50 Gy

80

V40 < 50%

80

Bladder

V45 < 35%

50

Femoral head (left)

V50 < 2%

50

Femoral head (right)

V50 < 2%

50

Body

Dmax ≤ 53.50 Gy

300

  1. Abbreviations: PTVo The planning target volume excluding the small bowel, PTVi The planning target volume inside the small bowel, Dmax The maximum dose received, Dmean The mean dose, Dn% Dose received by the n% volume of the target volume, Vx Percentage volume irradiated by x Gy or more of a certain structure