Skip to main content

Table 2 Dose-volumetric parameters of the PTV

From: Comparison of treatment plans between IMRT with MR-linac and VMAT for lung SABR

DV parameter

VMAT

IMRT with MR-linac

p

D98% (Gy)

59.1 ± 0.3

60.3 ± 1.1

<  0.001

D95% (Gy)

60.0 ± 0.0

60.7 ± 1.2

0.015

D5% (Gy)

64.9 ± 1.3

63.8 ± 1.6

0.026

D2% (Gy)

65.3 ± 1.4

64.2 ± 1.7

0.035

Minimum dose (Gy)

54.6 ± 1.4

58.7 ± 1.3

<  0.001

Maximum dose (Gy)

66.8 ± 1.3

63.8 ± 1.5

< 0.001

Mean dose (Gy)

63.0 ± 0.8

62.3 ± 1.4

0.047

PTV volume (cm3)

27.2 ± 23.5 (ITV + 5 mm)

6.3 ± 7.7 (GTV + 3 mm)

< 0.001

Homogeneity index

0.099 ± 0.023

0.062 ± 0.014

< 0.001

  1. Note: DV parameter dose-volumetric parameter, VMAT volumetric modulated arc therapy, IMRT intensity modulated radiation therapy, MR-linac magnetic resonance image linear accelerator, Dn% dose received by at least n% volume of the planning target volume, PTV planning target volume, ITV internal target volume, GTV gross tumor volume