Table 4 Values of PA resulting from the comparison of the dose obtained from the plan reconstructed from expected positions with the dose obtained from a plan in which the absolute value of position errors |ε| were scaled up to a maximum Σ

From: Treatment verification using Varian’s dynalog files in the Monte Carlo system PRIMO

PA [%]
Region (prostate) Σ[mm]
  0.33 () 2.0 3.0 4.0 5.0 10.0
  RMS[mm]
  0.02 0.14 0.21 0.28 0.34 0.68
PTV 1 100 99.1 98.6 98.2 97.7 95.5
PTV 2 100 99.0 98.6 98.1 97.7 95.4
PTV 3 100 99.1 98.6 98.2 97.8 95.6
PTV 4 100 99.0 98.5 98.0 97.6 95.3
PTV total 100 99.0 98.6 98.1 97.7 95.4
Rectum 100 98.8 98.1 97.6 97.0 94.3
Bladder 100 98.4 97.6 98.9 96.2 92.7
Body 100 98.8 98.3 97.7 97.2 94.7
Region (H&N) Σ[mm]
  0.55 () 2.0 3.0 4.0 5.0 10.0
  RMS[mm]
  0.03 0.10 0.14 0.19 0.24 0.47
PTV 1 100 99.5 99.1 98.8 98.5 97.0
PTV 2 100 99.5 99.3 99.1 98.8 97.6
Spinal canal 99.9 99.4 99.1 98.8 98.6 97.3
Parotid left 99.9 99.8 99.6 99.3 99.2 98.5
Parotid right 99.9 99.1 98.3 98.1 97.2 95.2
Body 100 99.5 99.2 99.0 98.7 97.4
  1. The RMS is also reported for each case. First column () corresponds to the unmodified dynalog files