Table 1 Results of the comparison of the dose obtained from the original plan with the dose obtained from the plan reconstructed from the expected positions

From: Treatment verification using Varian’s dynalog files in the Monte Carlo system PRIMO

Region

Low resolution

 

High resolution

 

PA [%]

GPR 2,1 [%]

 

PA [%]

GPR 2,1 [%]

Prostate

PTV 1

99.5

100

 

100

100

PTV 2

99.5

100

 

100

100

PTV 3

99.6

100

 

100

100

PTV 4

99.7

100

 

99.9

100

PTV total

99.6

100

 

99.9

100

Rectum

99.7

99.8

 

99.9

99.9

Bladder

99.8

100

 

99.7

100

Body

99.9

99.4

 

99.9

99.4

H&N

PTV 1

99.5

99.9

 

99.7

100

PTV 2

99.7

99.4

 

99.9

99.9

Spinal canal

99.7

99.9

 

99.7

99.8

Parotid left

99.7

99.8

 

99.6

99.9

Parotid right

99.4

99.8

 

100

99.8

Body

99.8

99.4

 

99.8

99.4

  1. The dose distribution of the original was obtained considering the control points included in the DICOM RTPLAN file (Low resolution) and obtained by increasing the number of control points (High resolution) by linear interpolation of the leaf positions. A good agreement of both dose distributions was obtained in both cases