Table 1 Results of the comparison of the dose obtained from the original plan with the dose obtained from the plan reconstructed from the expected positions

From: Treatment verification using Varian’s dynalog files in the Monte Carlo system PRIMO

Region Low resolution   High resolution
  PA [%] GPR 2,1 [%]   PA [%] GPR 2,1 [%]
Prostate
PTV 1 99.5 100   100 100
PTV 2 99.5 100   100 100
PTV 3 99.6 100   100 100
PTV 4 99.7 100   99.9 100
PTV total 99.6 100   99.9 100
Rectum 99.7 99.8   99.9 99.9
Bladder 99.8 100   99.7 100
Body 99.9 99.4   99.9 99.4
H&N
PTV 1 99.5 99.9   99.7 100
PTV 2 99.7 99.4   99.9 99.9
Spinal canal 99.7 99.9   99.7 99.8
Parotid left 99.7 99.8   99.6 99.9
Parotid right 99.4 99.8   100 99.8
Body 99.8 99.4   99.8 99.4
  1. The dose distribution of the original was obtained considering the control points included in the DICOM RTPLAN file (Low resolution) and obtained by increasing the number of control points (High resolution) by linear interpolation of the leaf positions. A good agreement of both dose distributions was obtained in both cases