Table 3 The comparison of the delivery parameters for VMAT only, primary hybrid, and modified hybrid plan

From: Evaluation of optimal treatment planning for radiotherapy of synchronous bilateral breast cancer including regional lymph node irradiation

  VMAT only Primary hybrid plan Modified hybrid plan p value
Mean (SD) Mean (SD) Mean (SD)
Beam on time (s) 115.3 (11.6) 73.5 (8.8) 77.5 (8.8) < 0.001
MU segment 795.1 (131) 827.5 (59.2) 811.7 (67.3) 0.424
  1. Abbreviations: MU monitor unit