Table 4 Summary of predicted per-organ 3D GPR derived from each TPS (XiO or RayStaion)

From: Comparison of gamma index based on dosimetric error and clinically relevant dose–volume index based on three-dimensional dose prediction in breast intensity-modulated radiation therapy

predicted per-organ 3D GPR

XiO

RayStation

P value

Mean ± SD (range)

Mean ± SD (range)

PTV

 3%/3 mm

94.9 ± 2.1

93.1 ± 2.4

p < 0.001

(89.2–98.2)

(88.2–96.8)

 3%/2 mm

93.9 ± 2.4

91.6 ± 2.8

p < 0.001

(88.0–97.8)

(86.4–96.2)

 2%/2 mm

91.9 ± 3.1

89.4 ± 5.2

Not Significant

(85.4–96.0)

(76.8–96.4)

Ipsilateral lung

 3%/3 mm

73.3 ± 9.1

85.9 ± 5.0

p < 0.001

(53.1–86.2)

(75.4–93.3)

 3%/2 mm

67.7 ± 8.7

84.6 ± 5.4

p < 0.001

(47.6–78.5)

(73.0–93.1)

 2%/2 mm

63.1 ± 8.9

84.3 ± 5.3

p < 0.001

(43.9–79.3)

(73.3–93.0)

Heart

 3%/3 mm

97.6 ± 3.9

98.8 ± 3.5

p < 0.05

(84.1–100)

(83.8–100)

 3%/2 mm

95.4 ± 3.8

98.4 ± 3.5

p < 0.001

(84.6–99.9)

(83.8–100)

 2%/2 mm

92.6 ± 4.8

98.4 ± 3.4

p < 0.001

(82.7–99.0)

(83.8–100)