Table 1 Description of 21 plans

From: Comparison of gamma index based on dosimetric error and clinically relevant dose–volume index based on three-dimensional dose prediction in breast intensity-modulated radiation therapy

 

Value (range)

# Beams

2 (all plans)

Minimum segment size

2 × 2 cm2 (all plans)

Minimum monitor units (MUs) per segment

5 MU (all plans)

Mean (range) # segments

35 ± 7 (19–46)

Mean (range) equivalent square side (cm)

8.6 ± 0.8 (7.4–10.2)

Dose–volume constraint of PTV: D95%

≥95% (47.5 Gy)

Dose–volume constraint of ipsi. lung: V20Gy

< 15% (per protcol), < 20% (variation acceptable)

Dose–volume constraint of heart: Dmean

< 4 Gy

  1. D95%: dose coverage 95% volume, ipsi.: ipsilateral, V20Gy: volume receiving at least 20 Gy, Dmean: dose coverage mean volume