Skip to main content

Table 4 Record on dosimetric endpoint value change with each stage ends for case 10

From: Voxel-based automatic multi-criteria optimization for intensity modulated radiation therapy

ROI

Endpoint

Original plan

Stage 1

Stage 2

Stage 3

Stage 4

PTV

D99%

42.76 Gy

42.92 Gy

42.89 Gy

42.91 Gy

42.86 Gy

PTV

D97%

43.7 Gy

43.7 Gy

43.7 Gy

43.7 Gy

43.7 Gy

PTV

V115%

0

0

0

0

0

PTV

V110%

0

0

0

0

0

Rectum

Dmax

46.1 Gy

47.3 Gy

47.5 Gy

47.5 Gy

49.7 Gy

Bladder

Dmax

46.5 Gy

47.5 Gy

47.7 Gy

47.9 Gy

48.1 Gy

Bowel

Dmax

46.5 Gy

47.5 Gy

47.7 Gy

47.9 Gy

47.7 Gy

Femoral heads

Dmax

36.5 Gy

38.9 Gy

39.7 Gy

39.9 Gy

38.3 Gy

Rectum

V45

21.5%

16.23%

15.54%

15.71%

17.43%

Bladder

V45

45.78%

30.07%

29.47%

30.01%

32.28%

Rectum

V30

93.51%

71.31%

71.43%

71.41%

61.30%

Bowel

V40

11.25%

11.07%

10.88%

10.95%

10.98%

Femoral heads

V30

6.71%

6.34%

6.88%

6.92%

5.75%

Bone

V20

73.80%

61.1%

60.82%

60.99%

60.66%

Bone

V10

88.80%

79.09%

78.71%

78.81%

78.12%