Skip to main content

Table 4 TRAK per seed (in μGym2) for different prescription depths (1 mm – 10 mm in 1 mm intervals) for standard COMS plaques loaded with 125I seeds (model IAI-125A)

From: A practical approach to estimating optic disc dose and macula dose without treatment planning in ocular brachytherapy using 125I COMS plaques

Prescription depth (mm)

Plaque size (mm) in diameter

10

12

14

16

18

20

22

1

299.9

226.3

161.2

191.6

134.8

132.3

157.3

2

424.9

306.2

212.2

241.0

168.0

162.0

194.2

3

587.2

406.6

274.4

301.3

206.3

195.8

234.6

4

788.9

530.2

349.5

373.9

251.2

234.7

280.1

5

1030.6

677.1

438.1

459.8

303.2

279.4

331.2

6

1311.2

848.0

540.9

559.7

363.2

330.7

389.3

7

1634.7

1045.2

659.3

675.4

432.6

389.9

455.7

8

2008.7

1274.1

797.3

810.4

513.4

458.6

532.3

9

2438.7

1536.9

955.4

965.1

605.8

537.1

619.8

10

2923.5

1833.6

1133.9

1140.1

710.4

626.0

718.6