Skip to main content

Table 3 Comparison of the mean pixel values in fat and muscle between the pCT, CBCT and corrected CBCT images with the three different DR algorithms

From: Assessing the impact of choosing different deformable registration algorithms on cone-beam CT enhancement by histogram matching

Mean pixel values [HU]

Patient’s number

Fat

pCT

CBCT

CBCTDR-NCC

CBCTDR-MI

CBCTDR-NMI

 1

− 108.16

− 144.37

−90.73

−92.50

−92.06

 2

−87.68

− 163.92

−84.98

−88.25

−85.77

 3

−104.83

− 139.75

−90.03

−91.48

−91.16

 4

− 103.80

− 151.60

−104.64

− 105.23

−105.19

 5

−103.43

− 170.34

− 102.28

−103.16

−102.89

 6

−109.43

− 130.55

−98.50

− 99.30

−99.09

 7

−97.70

− 148.56

−97.08

−98.47

−97.97

 8

−108.62

− 213.92

− 108.76

− 108.53

−108.04

 9

− 110.23

−160.35

− 105.87

− 106.76

−106.82

 10

−99.50

−123.33

−100.24

− 101.55

−101.40

RMSD

 

56.83

8.19

7.38

7.57

 

Muscle

 1

49.45

−1.37

31.76

30.62

30.68

 2

51.61

25.39

42.09

41.12

41.68

 3

47.59

59.00

53.25

54.53

53.25

 4

42.17

45.81

50.65

51.61

50.93

 5

48.87

11.70

47.66

47.66

47.65

 6

49.68

40.03

50.09

50.78

50.25

 7

49.73

28.39

45.44

46.12

46.01

 8

46.46

13.39

36.65

39.04

39.34

 9

48.85

56.30

52.18

51.95

51.86

 10

49.44

51.98

45.97

47.25

46.40

RMSD

 

25.49

7.72

8.28

8.02