From: Technical and dosimetric implications of respiratory induced density variations in a heterogeneous lung phantom

Fig. 7

A planner view of the dose distribution mismatch between AAA versus AXB computations. Plans were calculated for a AIP, b FB, c MIP, and d PWD data sets for target range of motion spanning from 1 to 3 cm (left to right). The analysis shown here is based on stringent gamma criteria (1%, 0 mm), where regions in red are indicating higher AAA dose calculation, versus blue for lower dose as with respect to AXB