From: Validation of new transmission detector transmission factors for online dosimetry: an experimental study
Energy
aIMRT
bVMAT
cSABR
6 MV
0.878 ± 0.002
0.877 ± 0.002
0.878 ± 0.006
6 dFFF
0.825 ± 0.003
0.823 ± 0.001
0.827 ± 0.007
10 MV
0.912 ± 0.003
0.914 ± 0.001
0.913 ± 0.009
10 FFF
0.883 ± 0.002
0.882 ± 0.003
0.885 ± 0.005