Fig. 3 | Radiation Oncology

Fig. 3

From: PRIMO Monte Carlo software benchmarked against a reference dosimetry dataset for 6 MV photon beams from Varian linacs

Fig. 3

Output factors for open fields at dmax reported by IROC–H, and calculated with PRIMO using the DPM algorithm. The maximum differences between simulations and IROC–H data are − 1.6% for Clinac 2100, and 0.4% for TrueBeam. Uncertainty bars show (with k=2) the standard deviation of the IROC–H data, and the statistical uncertainty of the simulations. Data points are artificially separated along the horizontal axis for clarity