Skip to main content
Fig. 1 | Radiation Oncology

Fig. 1

From: PRIMO Monte Carlo software benchmarked against a reference dosimetry dataset for 6 MV photon beams from Varian linacs

Fig. 1

Percentage depth–doses of a 10×10 cm2 field reported by IROC–H, and calculated with PRIMO using the DPM algorithm. The maximum differences between simulations and IROC–H data are 2.3% for Clinac 2100, and 0.6% for TrueBeam. Uncertainty bars show (with k=2) the standard deviation of the IROC–H data, and the statistical uncertainty of the simulations. For most data points, the bars are smaller than the symbol size. Data points are artificially separated along the horizontal axis for clarity

Back to article page