Table 3 Group systematic error (M), systematic error (Σ), and random error (σ) for intrafractional error of the prostate with OBI-FM and CBCT-ST for 206 post-treatment datasets

From: The potential failure risk of the cone-beam computed tomography-based planning target volume margin definition for prostate image-guided radiotherapy based on a prospective single-institutional hybrid analysis

 

AP /mm

SI /mm

LR /mm

OBI-FM for 1177 treatment including 971 and 206 treatments

M

0.56

0.22

0.34

Σ

0.68

0.76

0.26

σ

1.66

1.62

0.75

OBI-FM for 206 treatments

M

0.49

0.22

0.24

Σ

0.85

1.00

0.43

σ

1.47

1.66

0.75

CBCT-ST for 206 treatments

M

0.00

−0.04

0.12

Σ

0.68

0.66

0.25

σ

1.08

1.21

0.50

  1. kV kilo voltage, OBI on-board imaging; CBCT cone-beam computed tomography, FM fiducial marker matching, ST soft tissue matching, AP anterior to posterior, SI superior to inferior, LR left to right